|
Environmental conditions can
have an effect on sensor accuracy!
With
the increasing prevalence of gas detectors used daily for ensuring workers
safe entry into confined spaces, it is important that these instrument
users understand the factors which may influence the reliability of the
information produced by these monitoring tools. This article is intended
to briefly review some of the concerns regarding the measurement for explosive
atmospheres with combustible gas (LEL) indicating instruments.
There are three types of sensors
which are commonly used in confined space monitors; oxygen, combustible
gas (LEL), and toxic gas sensors. Each type of sensor uses a slightly
different detection principle. The kinds of conditions that can affect
accuracy vary from one type of sensor to the next. The type of sensor
which is most prone to being affected by the atmosphere in which it is
being used tends to be the combustible gas sensor. Age and usage can have
a serious effect on sensitivity.
Chronic exposure to silicone
containing substances (found in many lubricants), the tetra-ethyl-lead
found in "leaded " gasoline, halogenated hydrocarbons (FreonsŪ,
or solvents such as trichloroethylene and methylene chloride), high concentrations
of hydrogen sulfide or even very high concentrations of combustible gas
may lead to degraded combustible sensor performance. In most cases all
this means is that the sensitivity is adjusted upwards at the time the
instrument is calibrated. In the worst case, the sensor may need to replaced.
Once again, verifying the accuracy of the sensors on a regular basis is
essential to assuring worker safety.
How combustible sensors work
The minimum amount of a combustible
gas or vapor in air which will explosively burn if a source of ignition
is present is the Lower Explosive Limit (LEL) concentration. Combustible
gas readings are given in percent LEL, with a range of zero to one-hundred
percent explosive. Combustible sensors contain two coils of fine wire
coated with a ceramic material to form beads. The "active" bead
is coated with a palladium based material that allows catalyzed combustion
to occur on the surface of the bead. The "reference" bead lacks
the catalyst coating, but in other respects exactly resembles the active
bead.
Any combustible vapors which
are present will be subject to catalytic combustion on the surface of
the active bead, heating this bead to a higher temperature. The temperature
of the untreated reference bead is unaffected by the presence of gas.
The difference between the temperatures of the two beads is proportional
to the amount of combustible gas present. Since the beads are strung on
the opposite arms of a Wheatstone Bridge electrical circuit, the instrument
perceives this as a change in the electrical resistance in the circuit.
It is this change in resistance due to differential heating that is used
by the instrument to provide a reading.
Combustible gas sensors are
non-specific and respond to all combustible gases and vapors. It is not
necessary for the combustible vapor to be present in LEL concentrations.
Even trace amounts of combustible gas can be detected by this method.
The amount of heat produced by the combustion of a particular gas on the
active bead will reflect the "Heat of Combustion" for that gas.
Heats of combustion may vary from one combustible gas to another. For
this reason readings may vary between equivalent concentrations of different
combustible gases.
A combustible gas and vapor
reading instrument may be calibrated to any number of different gases
or vapors. If an instrument is only going to be used for a single type
of gas over and over again, it is usually best to calibrate the instrument
to that particular hazard. As long as the gas that is encountered is the
same gas that was used during calibration, the readings will be exactly
accurate (to the tolerances of the instrument design).
Figure 1.0 illustrates what
may be seen when a combustible sensor is used to monitor gases other than
the one to which it was calibrated. The chart shows the "relative
response curves" of the instrument to several different gases.
Figure 1.0: Relative response
curves
Note that the response to
the gas to which the instrument was calibrated, the "calibration
standard", is still precisely accurate. For the other gases the responses
are a little off. In the case of some gases the readings are a little
high. This results in the instrument going into alarm a little bit early.
This type of error is not usually dangerous, since it simply results in
workers exiting the affected area sooner than they otherwise would have.
Gases which produce lower
relative readings than the calibration standard can result in a more potentially
dangerous sort of error. One way to reduce the potential for this type
of error is to use a lower alarm setting. It may be seen from the graph
that the amount of relative error decreases the lower the alarm point
is set. If the alarm point is set at 10 percent LEL, the differences due
to relative response of the combustible sensor are minimal.
Choosing the right calibration
mixture
The other method for reducing
the effects of this sort of error is in the choice of the calibration
gas used to calibrate the combustible sensor. The best results are obtained
when calibration is done using the same gas that is expected to be encountered
while actually using the instrument. When it is not possible to calibrate
directly to the gas to be measured, or when the combustible gas is an
unknown, a mixture which provides a sensor response which is more typical
of the range of combustible gases and vapors that will be encountered
should be selected.
Summation
Remember that in actual practice,
the relative response varies somewhat from sensor to sensor. Response
ratios may also shift over the life of a particular sensor. Most importantly,
if sensitivity is lost due to poisoning, it is frequently lost first with
regards to methane.
In most cases the loss of
occurs a little at a time. In some cases, however, the loss of sensitivity
can be almost immediate. This is the reason that reputable gas detector
manufacturers place so much emphasis on calibration, and why use of a
certified calibration gas is a good idea.
Most gas monitoring device
manufacturers recommend that the accuracy of any gas detector be verified
before each day's use. In addition, accuracy should be re-checked immediately
after any known exposure to contaminants.
Article Contributed by:
Timothy J. Kearney,
CSP
Vice President, Argus Group
Article Provided
by: The Argus Group
15075 E. Eleven Mile Road
Roseville, MI 48066
(810) 774-8900
Fax: (810) 774-2780
Toll Free 1-800-873-0456
Copyright
1998 - 2001 Argus Group - All rights reserved.
Last Update:
July 23, 2002
|