|
Go Back to Index
Environmental
conditions can have an effect on sensor accuracy!
With the increasing
prevalence of gas detectors used daily for ensuring workers safe entry
into confined spaces, it is important that these instrument users understand
the factors which may influence the reliability of the information produced
by these monitoring tools. This article is intended to briefly review
some of the concerns regarding the measurement for explosive atmospheres
with combustible gas (LEL) indicating instruments.
There are three types
of sensors which are commonly used in confined space monitors; oxygen,
combustible gas (LEL), and toxic gas sensors. Each type of sensor uses
a slightly different detection principle. The kinds of conditions that
can affect accuracy vary from one type of sensor to the next. The type
of sensor which is most prone to being affected by the atmosphere in which
it is being used tends to be the combustible gas sensor. Age and usage
can have a serious effect on sensitivity.
Chronic exposure
to silicone containing substances (found in many lubricants), the tetra-ethyl-lead
found in "leaded " gasoline, halogenated hydrocarbons (FreonsŪ,
or solvents such as trichloroethylene and methylene chloride), high concentrations
of hydrogen sulfide or even very high concentrations of combustible gas
may lead to degraded combustible sensor performance. In most cases all
this means is that the sensitivity is adjusted upwards at the time the
instrument is calibrated. In the worst case, the sensor may need to replaced.
Once again, verifying the accuracy of the sensors on a regular basis is
essential to assuring worker safety.
How
combustible sensors work
The minimum amount
of a combustible gas or vapor in air which will explosively burn if a
source of ignition is present is the Lower Explosive Limit (LEL) concentration.
Combustible gas readings are given in percent LEL, with a range of zero
to one-hundred percent explosive. Combustible sensors contain two coils
of fine wire coated with a ceramic material to form beads. The "active"
bead is coated with a palladium based material that allows catalyzed combustion
to occur on the surface of the bead. The "reference" bead lacks
the catalyst coating, but in other respects exactly resembles the active
bead.
Any combustible vapors
which are present will be subject to catalytic combustion on the surface
of the active bead, heating this bead to a higher temperature. The temperature
of the untreated reference bead is unaffected by the presence of gas.
The difference between the temperatures of the two beads is proportional
to the amount of combustible gas present. Since the beads are strung on
the opposite arms of a Wheatstone Bridge electrical circuit, the instrument
perceives this as a change in the electrical resistance in the circuit.
It is this change in resistance due to differential heating that is used
by the instrument to provide a reading.
Combustible gas sensors
are non-specific and respond to all combustible gases and vapors. It is
not necessary for the combustible vapor to be present in LEL concentrations.
Even trace amounts of combustible gas can be detected by this method.
The amount of heat produced by the combustion of a particular gas on the
active bead will reflect the "Heat of Combustion" for that gas.
Heats of combustion may vary from one combustible gas to another. For
this reason readings may vary between equivalent concentrations of different
combustible gases.
A combustible gas
and vapor reading instrument may be calibrated to any number of different
gases or vapors. If an instrument is only going to be used for a single
type of gas over and over again, it is usually best to calibrate the instrument
to that particular hazard. As long as the gas that is encountered is the
same gas that was used during calibration, the readings will be exactly
accurate (to the tolerances of the instrument design).
Figure 1.0 illustrates
what may be seen when a combustible sensor is used to monitor gases other
than the one to which it was calibrated. The chart shows the "relative
response curves" of the instrument to several different gases.
Figure 1.0: Relative response curves
Note that the response
to the gas to which the instrument was calibrated, the "calibration
standard", is still precisely accurate. For the other gases the responses
are a little off. In the case of some gases the readings are a little
high. This results in the instrument going into alarm a little bit early.
This type of error is not usually dangerous, since it simply results in
workers exiting the affected area sooner than they otherwise would have.
Gases which produce
lower relative readings than the calibration standard can result in a
more potentially dangerous sort of error. One way to reduce the potential
for this type of error is to use a lower alarm setting. It may be seen
from the graph that the amount of relative error decreases the lower the
alarm point is set. If the alarm point is set at 10 percent LEL, the differences
due to relative response of the combustible sensor are minimal.
Choosing
the right calibration mixture
The other method for
reducing the effects of this sort of error is in the choice of the calibration
gas used to calibrate the combustible sensor. The best results are obtained
when calibration is done using the same gas that is expected to be encountered
while actually using the instrument. When it is not possible to calibrate
directly to the gas to be measured, or when the combustible gas is an
unknown, a mixture which provides a sensor response which is more typical
of the range of combustible gases and vapors that will be encountered
should be selected.
Summation
Remember that in actual
practice, the relative response varies somewhat from sensor to sensor.
Response ratios may also shift over the life of a particular sensor. Most
importantly, if sensitivity is lost due to poisoning, it is frequently
lost first with regards to methane.
In most cases the
loss of occurs a little at a time. In some cases, however, the loss of
sensitivity can be almost immediate. This is the reason that reputable
gas detector manufacturers place so much emphasis on calibration, and
why use of a certified calibration gas is a good idea.
Most gas monitoring
device manufacturers recommend that the accuracy of any gas detector be
verified before each day's use. In addition, accuracy should be re-checked
immediately after any known exposure to contaminants.
Article Contributed
by:
Timothy J. Kearney,
CSP
Vice President, Argus Group
Article
Provided by: The Argus Group
46400 Continental
Drive
Chesterfield Twp., MI 48047
(586) 840-3200
Fax: (586) 774-2780
Toll Free 1-800-873-0456

|